Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Original Article
  • Published:

Reflectance Pulse Oximetry from Core Body in Neonates and Infants: Comparison to Arterial Blood Oxygen Saturation and to Transmission Pulse Oximetry

Abstract

OBJECTIVE: To compare pulse oximetry oxygen saturation (SpO2) measured by a novel reflectance method from core body to arterial oxygen saturation (SaO2) in neonates and infants. Transmission pulse oximetry (TPO) was measured for comparison.

STUDY DESIGN: We monitored 18 infants by the two pulse oximeters simultaneously. The reflectance pulse oximetry (RPO) (PRO2, ConMed, Utica, NY) was measured on the upper back or chest, while the TPO (N395-Nellcor, Pleasanton, CA) was measured from the finger of the infant on the left hand or feet. Data from the two methods were compared to functional SaO2 derived from blood sample drawn from arterial line for patient care and measured by a Co-oximeter (Ilex, Instrument Lab. Lexington, MA). The potential advantage of the RPO is demonstrated in a case of a premature infant with hypovolemic shock, where SaO2 or TPO could not be obtained but oximetry was available from the RPO.

RESULTS: We used for analysis 56 RPO and 32 TPO measurements. SpO2 obtained from the RPO was 88.3±9.8%, from the TPO 84.2±10.1%, and functional SaO2 was 88.2±11.7%, with correlation coefficient of 0.93 and 0.88, respectively (p<0.0001). The mean difference (bias) and standard deviation of the differences (precision) between the RPO and the TPO compared to functional SaO2 were −0.09±4.5% and 1.26±5.9% and the absolute errors were 3.2±3.1%, and 4.4±4.0%, respectively. The accuracy of both RPO and TPO was diminished when SaO2 was <85%, but only the RPO remained correlated with the functional SaO2.

CONCLUSIONS: Reflectance pulse oximetry measured from core body of neonates and infants is accurate and reliable and is comparable to the transmission SpO2 when compared to functional SaO2. We speculate that the reflectance method might be advantageous in cases of poor peripheral perfusion in neonates and infants.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Figure 1
Figure 2
Figure 3
Figure 4

Similar content being viewed by others

References

  1. Mower WR, Sachs C, Nicklin EL, Baraff LJ . Pulse oximetry as a fifth pediatric vital sign. Pediatrics 1997;99:681–686.

    Article  CAS  Google Scholar 

  2. Katzman GH . The newborn's SpO2: a routine vital sign whose time has come ? Pediatrics 1995;95:161–162.

    CAS  PubMed  Google Scholar 

  3. Sinex JE . Pulse oximetry: principles and limitations. Am J Emerg Med 1999;17:59–67.

    Article  CAS  Google Scholar 

  4. Hay WW, Brockway JM, Eyzaguirre M . Neonatal pulse oximetry: accuracy and reliability. Pediatrics 1989;83:717–722.

    PubMed  Google Scholar 

  5. Durand M, Ramanathan R . Pulse oximetry for continuous oxygen monitoring in sick newborn infants. J Pediatr 1986;109:1052–1056.

    Article  CAS  Google Scholar 

  6. Gehring H, Hornberger C, Matz H, Konecny E, Schmuker P . The effect of motion artifact and low perfusion on the performance of new generation of pulse oximeters in volunteers undergoing hypoxemia. Respir Care 2002;47:48–60.

    PubMed  Google Scholar 

  7. Poets CF, Southall DP . Noninvasive monitoring of oxygenation in infants and children: practical considerations and areas of concern. Pediatrics 1995;95:161–162.

    Google Scholar 

  8. Faisst K, Hannon W, Jorgensen JS, et al. Reflectance pulse oximetry in neonates. Eur J Obstet Gynecol Reprod Biol 1995;61:117–122.

    Article  CAS  Google Scholar 

  9. Fanconi S, Tschupp A . Accuracy of a new transmittance–reflectance pulse oximetry sensor in critically ill neonates. Crit Care Med 1994;22:1142–1146.

    Article  CAS  Google Scholar 

  10. Takatani S, Davies C, Sakakibara N, et al. Experimental and clinical evaluation of a noninvasive reflectance pulse oximeter sensor. J Clin Monit 1992;8:257–266.

    Article  CAS  Google Scholar 

  11. Thilo EH, Anderson D, Wasserstein ML, Schmidt J, Luckey D . Saturation by pulse oximetry: comparison of the results obtained by instruments of different brands. J Pediatr 1993;122:620–626.

    Article  CAS  Google Scholar 

  12. Wukitsch MW, Petterson MT, Tobler DR, Pologe JA . Pulse oximetry: analysis of theory, technology, and practice. J Clin Monit 1988;4:290–301.

    Article  CAS  Google Scholar 

  13. Ramanathan R, Durand M, Larrazabal C . Pulse oximetry in very low birth weight infants with acute and chronic lung disease. Pediatrics 1987;79:612–617.

    CAS  PubMed  Google Scholar 

  14. Pologe JA, Raney DM . Effects of fetal hemoglobin on pulse oximetry. J Perinatol 1987;7:324–326.

    CAS  PubMed  Google Scholar 

  15. Rajadurai VS, Walker AM, Yu Vy, Oates A . Effect of fetal haemoglobin on the accuracy of pulse oximetry in preterm infants. J Paediatr Child Health 1992;28:43–46.

    Article  CAS  Google Scholar 

  16. Sardesai S, Durand M, MacEvoy C, Johnson C, Maarek JM . Pulse oximetry in newborn infants with birth weights of 620-4285 grams receiving dopamine and dobutamine. J Perinatol 1996;16:31–34.

    CAS  PubMed  Google Scholar 

  17. Hay Jr WW, Rodden DJ, Collins SM, Melara DL, Hale KA, Fashaw LM . Reliability of conventional and new pulse oximetry in neonatal patients. J Perinatol 2002;22:360–366.

    Article  Google Scholar 

  18. Fanconi S . Reliability of pulse oximetry in hypoxic infants. Pediatrics 1988;122:424–427.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Data were presented in part as a poster at the Pediatric Academic Society meeting in Baltimore 2002 (“late breaker” session) and in the ATS meeting in Seattle 2003. This work was supported by ConMed Corporation, Utica, NY, USA.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kugelman, A., Wasserman, Y., Mor, F. et al. Reflectance Pulse Oximetry from Core Body in Neonates and Infants: Comparison to Arterial Blood Oxygen Saturation and to Transmission Pulse Oximetry. J Perinatol 24, 366–371 (2004). https://doi.org/10.1038/sj.jp.7211102

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/sj.jp.7211102

This article is cited by

Search

Quick links