Abstract
Electron pumps generate a macroscopic electric current by controlled manipulation of single electrons. Despite intensive research towards a quantum current standard over the last 25 years, making a fast and accurate quantized electron pump has proved extremely difficult. Here we demonstrate that the accuracy of a semiconductor quantum dot pump can be dramatically improved by using specially designed gate drive waveforms. Our pump can generate a current of up to 150 pA, corresponding to almost a billion electrons per second, with an experimentally demonstrated current accuracy better than 1.2 parts per million (p.p.m.) and strong evidence, based on fitting data to a model, that the true accuracy is approaching 0.01 p.p.m. This type of pump is a promising candidate for further development as a realization of the SI base unit ampere, following a redefinition of the ampere in terms of a fixed value of the elementary charge.
Similar content being viewed by others
Introduction
The AC Josephson effect and quantum Hall effect have revolutionized electrical metrology by providing stable reference standards linked to fundamental physics rather than particular artefacts1. The resulting system of electrical units is coupled to the SI only through very difficult and time-consuming electro-mechanical experiments2, and recently a debate has ignited about redefining the SI ampere in terms of the fundamental elementary charge e3. A redefined ampere could be realised directly using a quantized electron pump, which generates a current by moving single electrons rapidly and in a controlled way. The electron pump accepts a periodic input signal at a repetition frequency f, and transports an integer number n of electrons between source and drain leads for each cycle of the input to yield a current IP=nef. Application as a quantum standard of electric current requires a pump having a combination of accuracy, simplicity of operation and ability to generate a reasonably high current IP≥100 pA4. In the past two decades, many pump technologies have been investigated in pursuit of this goal. These include chains of submicron normal-metal tunnel junctions5,6, quantum dots driven by a surface acoustic wave7 and normal metal/superconductor turnstiles8,9. Although 0.015 parts per million (p.p.m.) accuracy has been demonstrated for IP≈1 pA5, and IP≈500 pA with around 100 p.p.m. accuracy10, none of these technologies have yet demonstrated the required combination of accuracy and large enough current output.
Our tunable-barrier electron pump (scanning electron microscope (SEM) image in Fig. 1a, with associated biasing voltages also shown) consists of a conducting channel etched in a two-dimensional electron gas (2-DEG), crossed by two metallic gates, which we denote the entrance and exit gates. Potential barriers are created in the 2-DEG below the gates by applying negative voltages VG1 and VG2 to these gates. An isolated quantum dot, holding a small number of electrons, forms between the gates. A drive signal VRF(t) superimposed on VG1 periodically lowers the entrance barrier below the Fermi level of the source lead, picking up electrons and lifting them over the exit barrier into the drain lead. The exit barrier is kept high (>10 meV above the Fermi level) to suppress unwanted thermally activated and cotunnel transport11. A series of schematic potential diagrams in Fig. 1b illustrate the pump cycle for the case of n=1. The pump operates at zero source-drain bias voltage; the direction of the pump current is determined by which gate is driven with the AC signal. If the amplitude of VRF is large enough, the probability of the trapped electron being ejected into the drain lead (frame 4 of Fig. 1b) is unity, and the current quantization mainly depends on the initial part of the pump cycle (frame 2). Here, the relatively large number of electrons initially trapped by the rising entrance barrier is reduced to just one by a cascade of 'back-tunnel' events whereby excess electrons return to the source lead (red arrow)12,13. Owing to the Coulomb charging energy of the dot, the back-tunnel probability for the last electron is many orders of magnitude less than for the second electron, so the convergence on the 1-electron state is robust and insensitive to details of the potential barrier shape. The GaAs pump has an important advantage compared with other types of electron pump, which is that the quantization accuracy can be improved with an external tuning parameter, the perpendicular magnetic field B14,15,16,17,18. The mechanism for this improvement is not fully understood, but the additional confinement of electrons within the dot owing to the magnetic field has an important role18.
Previous high-resolution measurements19,20 used sine wave drive for VRF(t). The sine wave cycle is shown in Fig. 1c with numbered points approximately indicating the stages of the pump cycle shown in Fig. 1b. These measurements could not be extended above f≈350 MHz because the quantized plateau degraded with further increase in f 20. The general mechanism for this is not currently understood, although clear signatures of back-tunneling due to nonadiabatic excitation at high f have been seen in some samples21. To extend the pump operation range to higher frequencies, we developed a technique using an arbitrary waveform generator (AWG) to generate a custom VG1(t) waveform that takes into account the time-dependent electron tunelling dynamics. To design the waveform, first we note, as shown schematically in Fig. 1c, that, for sine wave drive, most of the cycle time is used to raise the trapped electron over the exit barrier, and following ejection, lower the empty dot in readiness for the next cycle. This is a consequence of the height of the exit barrier and the large amplitude (≈0.5 V) of the entrance barrier modulation. Second, we postulate that the increased error rate at high frequency is related to the rate at which the entrance barrier is raised during the early part of the pump cycle, when back-tunneling occurs. Our AWG waveform, shown in Fig. 1d, was designed with a small initial dVG1/dt, speeding up for subsequent parts of the pump cycle. Our technique has some similarity to earlier work employing trapezoidal gate pulses with variable rise time12, and we will return to this point later in the Discussion section. The output of the AWG measured through a similar transmission line to the one used in the experiment showed a slight smearing of the actual VG1(t) applied to the entrance gate (red line in Fig. 1d), but this does not affect the important part of the cycle, which contains only low-frequency components. The 'slow' part of the waveform with repetition frequency f is approximately a segment of sine wave of frequency f/5 (dotted line), enabling us to slow down the electron capture process by a factor of 5. Thus, we expect the pump driven with the AWG pulse waveform at frequency f to have the same characteristics as it would have with sine wave drive at frequency f/5.
Using the AWG waveform, and employing a more accurate current measurement system (described in detail in the Methods section), we demonstrate a semiconductor pump generating currents up to 150 pA accurate to better than 1.2 p.p.m., the limit set by our measurement uncertainty. Theoretical analysis predicts that the true pump accuracy is between one and two orders of magnitude better than our experimental uncertainty, making this type of pump an extremely promising candidate for development as a future quantum representation of the ampere.
Results
Accuracy of the pump
We cooled our pumps to 300 mK and measured the pump current IP using a technique described in more detail in the Methods section. All high-resolution data in Figs 2,3,4 and 6 were obtained in a perpendicular magnetic field B=14 T. Most measurements reported in this paper were obtained on a single sample, denoted sample 1. Some measurements were made on an additional sample, denoted sample 2. Data are for sample 1 unless stated otherwise. Figure 2 shows a measurement of the pump current on the n=1 plateau as the exit gate voltage is varied. The 9 red data points were obtained in three experimental runs spread over 4 different days, and their mean and s.d. are shown by the horizontal pink line (with error bar indicating the 1.2 p.p.m. systematic uncertainty) and grey box, respectively. This data demonstrate the stability, plateau flatness and absolute accuracy of the pump at the 1 p.p.m. level. The accuracy can be seen by comparing the mean, with its associated uncertainty, with the value of ef indicated by the blue dotted line. We used the 2006 CODATA value of e, which has an uncertainty, insignificant for this measurement, of 0.025 p.p.m.22. On the plateau, the difference IP−ef is (−0.077±0.18) fA or (−0.51±1.2) p.p.m.
Quantization improvement using AWG drive
The dramatic effect of using AWG pulse drive on sample 1 is shown in more detail in Fig. 3. First, we made high-resolution measurements of the pump current as a function of VG2, using conventional sine wave drive, shown in Fig. 3a. The quantity plotted is the fractional deviation of pump current from ef in p.p.m., defined as ΔIP≡106(IP−ef)/ef. At 400 MHz, a broad plateau is seen, accurately quantized within the uncertainty of ~3 p.p.m., which is dominated by the random uncertainty UR for this combination of f and averaging time. When the frequency is increased to 630 MHz, the plateau becomes narrower, although it remains quantized over a finite range of VG2. Further increase in f results in a catastrophic loss of quantization. Figure 3b and c illustrate how the quantization can be recovered by switching to the AWG pulse waveform illustrated in Fig. 1d. At 630 MHz, the pulse waveform results in a ≈×2 increase in the width of the plateau compared with the sine wave drive, but at 945 MHz, where no plateau was present on this expanded current scale with sine wave drive, the pulse waveform completely restores accurate quantized pumping. Data for sample 2, at f=630 MHz, is shown in Fig. 4a. This sample did not exhibit as good a quantized plateau as sample 1, illustrated by the lack of a plateau using sine wave drive. However, the plateau could be progressively recovered using AWG waveforms with slower rising edges, illustrated in Fig. 4b. Waveform AWG1 (AWG2) had a rising edge corresponding to a sine wave of frequency 315 (218) MHz.
Fitting to the decay cascade model
The improvement using the AWG pulse drive can be demonstrated more quantitatively by considering theoretical predictions for the overall shape of the IP(VG2) plateau. In Fig. 5a, we plot the measured average number of electrons pumped per cycle n≡IP/ef, for four combinations of magnetic field and pumping frequency, using sine wave drive. In contrast to the high-resolution data of Figs 2,3 and 4, these data were taken over a wide range of VG2, with relatively low current resolution, to show the full transition n=0→1. As B is increased, the n=1 plateau becomes wider and the transitions between plateaux become sharper, resulting in improved current accuracy on the plateau, while increasing f has the opposite effect. The n(VG2) data are fitted to an analytical formula derived from the decay cascade model12,13:
where a, Δ1 and Δ2 are fitting parameters. The exponential approach to the n=1 quantized plateau, as VG2 is made more positive, is understood as an exponentially decreasing probability of the electron tunnelling back out of the dot into the source lead, while excess electrons tunnel back out with essentially unit probability (Fig. 1b). The parameter δ2=Δ2−Δ1 can be used as a practical figure of merit for evaluating the accuracy of a given pump, because there is a simple relationship between δ2 and the predicted pump error ≡1−nFIT at the point of inflection in IP(VG2)13. The point of inflection can be found from a relatively fast, low-resolution measurement. The fit line and its derivative for the 1-GHz data are shown in Fig. 5b with axes expanded to show .
Comparison of data with decay cascade model
Our high-resolution measurements enable us to investigate the agreement between efnFIT, obtained from low-resolution data, and the actual value of the pump current on the plateau. Figure 5c and d show a fit to low-resolution data, and the comparison with the high-resolution data, using AWG drive at 945 MHz. It can be seen that the fit slightly underestimates the width of the plateau, and consequently is an overestimate of the error at VG2=VG2,MIN, consistent with earlier, lower-resolution, measurements20. The other high-resolution data sets contributing to our assessment of the pump accuracy posessed similar properties; the fit always underestimated the true plateau flatness. Consequently, we consider as a lower limit (in other words, a pessimistic estimate) to the predicted pump accuracy, which for the 945 MHz data is approaching 0.01 p.p.m.
In Fig. 5e, we show δ2 as a function of B deduced from the experimental data measured with sine wave drive at f=150 MHz (left panel) and as a function of f measured at B=14 T with sine drive (right panel, open circles) and AWG pulse drive (right panel, filled circle). Horizontal lines show the thresholds for obtaining predicted accuracies of 1, 0.1 and 0.01 p.p.m. For f≈1 GHz using pulse AWG drive and B=14 T, we predict an accuracy better than 0.1 p.p.m. The value of δ2 extracted from the AWG pulse data at f=945 MHz roughly matches that for f~200 MHz using sine drive. This is in good agreement with the ×5 scaling of the time scale expected from the design of the pulse waveform. The data of Fig. 5e together with Figs 3 and 4 support our assumption that the pump errors at high frequency originate during the first part of the pump cycle, and can be minimized by slowing down this part of the cycle at the expense of other parts.
Summary of pump accuracy
In Fig. 6, we summarize the main experimental finding of this study. This figure shows the pump current averaged over the VG2 plateau for seven experimental runs at three frequencies for sample 1, and a single run for sample 2. The total measurement time is around 10 h per data point. We see that any deviation of the pump current from ef is less than our total measurement uncertainty , which is as low as 1.2 p.p.m. for the highest pump frequencies where the larger signal-to-noise ratio reduces the contribution of UR to a negligible level. A full investigation of the invariance of the pump current as a function of the additional control parameters VG1, the source-drain bias voltage and the amplitude of VRF is beyond the scope of this study, but an initial investigation into the invariance of IP(VG1) was undertaken. The data point in Fig. 6 marked with a star was obtained with VG1 shifted by 10 mV away from the plateau centre. No significant deviation from ef was seen following this adjustment.
Discussion
We have demonstrated that the pumping errors at high frequency can be reduced by many orders of magnitude by the use of specially designed gate drive waveforms with a slow initial rising edge. The data of Figs 3 and 4 support our postulate that errors at high pumping frequency arise during the initial phase of the pump cycle, when the number of electrons in the dot is reduced by the back-tunnel cascade. However, the reason for this is unclear. Indeed, the back-tunnel model12,13 does not predict any characteristic time-scale for the decay cascade as a consequence of assuming a strictly exponential dependence for the tunnel rates on VG1. Instead, it predicts a shift of the plateau to more negative VG2 as the rise time of VG1 is decreased. As already noted, one experiment on silicon MOSFET pumps provides support for the simple back-tunnel model: as the rise time of trapezoidal gate drive pulses was decreased from 20 ns to 2 ns, the plateaus shifted to more negative values of the top gate that controlled the dot depth, but the plateau shape did not change12. We do not believe that this result contradicts our own findings. We have observed the effect of decreasing rise time through the degredation of initially very flat plateaus measured to high resolution. In contrast, the plateaus presented in the variable rise-time experiment of ref. 12 were clearly not flat even at the longest rise time investigated. It is likely that the plateau flatness in this case was not limited by the rise time, and if much shorter rise times than 2 ns had been employed, a further reduction in the plateau flatness would have been seen, in agreement with our experiments. Understanding the upper frequency limit of accurate pumping in these devices requires more accurate physical models of the pumping mechanism, for example by incorporating a physically realistic model of the time dependence of tunnel rates into the back-tunnel model, or including the effect of nonadiabatic excitations21.
This is the first time an electron pump current has been directly compared with a reference current generated outside the pump cryostat with an uncertainty of order 1 p.p.m. In experiments using metallic pumps, a small pump current pA was compared with units of voltage, capacitance and time by charging a capacitor located in the same cryostat as the pump, and measuring the resulting voltage change across the capacitor6,23. The lowest uncertainty in these experiments was 0.92 p.p.m.24, similar to that in the present work, but with a current 300 times smaller. At the 100-pA current level of the semiconductor pump, a direct realization of the quantum metrological triangle25 at the 10−7 uncertainty level is feasible within an averaging time of several hours. Direct calibrations of a pico-ammeter or a micro-ammeter using a suitable high-gain current comparator26,27,28 are also possible. This makes our result a significant step towards the application of electron pumps in primary electrical metrology. We have shown that the pump current is invariant in one control paramter (VG2) and equal to ef within 1.2 p.p.m. Future experiments will investigate in detail the invariance of the current across the additional control parameters and over a range of samples.
Another class of experiment uses mesoscopic charge detectors capactively coupled to part of the pump circuit to count transport errors at the single electron level5. These experiments enable measurements of much smaller error rates than are possible by averaged current measurements such as those reported in this paper. Furthermore, they provide a direct measure of the pump error that is competely independent of the performance of any reference standard. If the electron pump is to be used as the current leg of a metrological triangle experiment, charge-counting accuracy tests are important because they allow the experiment to distinguish between errors in the number of transported charge quanta, and a more fundamental correction to the value of the charge on each quantum29. Error-counting experiments at the GHz frequencies employed in semiconductor pumps are technically challenging owing to the limited bandwidth of detectors, but some initial progress has already been made30,31, and these experiments will continue to be persued.
In summary, we have demonstrated a milestone in single-electron control, by transferring ≈109 electrons per second through a semiconductor quantum dot with a resulting current accuracy better than 1.2 p.p.m. By using shaped pulses to drive the control gate, we are able to limit errors and operate the pump accurately at higher frequency than was previously possible. These results are extremely encouraging for the development of a quantum current standard, redefinition of the unit ampere and more generally for precise high-speed control of electrons in semiconductor devices.
Methods
Sample and cryostat
We fabricated our pumps on GaAs/AlxGa1−xAs wafers using standard techniques: wet etching to define the channel, and electron-beam lithography for the metallic gates19. The samples were cooled in a sorbtion-pumped helium-3 refrigerator with a base temperature of 300 mK. The special gate pulses were generated using a 12 GS/s Tektronix 7122B AWG.
Current measurement system
Conventional electronic pico-ammeters are limited in practice to ≈100 p.p.m. accuracy, even following a calibration, owing to drift in the gain of the pre-amp stage. To achieve p.p.m. accuracy, we generated a reference current IR with opposite polarity to IP by applying a voltage VC across a temperature-controlled (26.57±0.02) 1-GΩ standard resistor R (circuit in Fig. 7a). A room-temperature current pre-amplifier, with transimpedance gain ≈1010 V/A calibrated to 0.1% accuracy, measured the small difference between the pump and reference currents. To remove offset currents and thermoelectric potentials, VC and VRF were synchronously switched on and off with a cycle time of 150 s. The differences in current and voltage, ΔI and ΔV, respectively, were recorded for each cycle as indicated in the section of raw data shown in Fig. 7b. The pump current is given by IP=ΔV/R+ΔI+(r/R)ΔI, where r≈10 kΩ is the input resistance of the pre-amp. As , the final term can be neglected and the pump current is determined by the voltmeter reading, the value of R and the small residual current ΔI with the pre-amp gain calibration contributing a small (<0.1 p.p.m.) correction. The voltmeter and resistor were calibrated, via some intermediate steps, against the AC Josephson effect and quantum Hall resistance, respectively. Formally, we have measured current in 1990 units, using the agreed values of the von Klitzing and Josephson constants RK−90 and KJ−90. The uncertainties of RK and KJ in SI units are sufficiently small, less than 0.1 p.p.m., that we can quote our result in SI units without a significant increase in uncertainty. The resistor was calibrated at a voltage of 10 V, with additional calibrations performed at higher voltages to check for possible power-coefficient effects as shown in Supplementary Fig. S1. No power coefficient was found in the voltage range 10 V ≤V≤100 V. Below 10 V, the resistor dissipates <100 nW and power coeffiecient effects are expected to be insignificant.
Uncertainty
The total systematic uncertainty US in our measurement of the pump current is 1.2 p.p.m., dominated by the 0.8-p.p.m. systematic uncertainty in the calibration of the resistor. The 2σ calibration uncertainty quoted for 1 GΩ resistors at NPL is 1.6 p.p.m. The relevent entry on the peer-reviewed calibration and measurement capability data base is available at http://kcdb.bipm.org/AppendixC/country_list_search.asp? CountSelected=GB&iservice=EM/DC.2.1.3. All unceratinties quoted in this paper are 1σ and have been rounded to the nearest 0.1 p.p.m. A full break down of the uncertainty is shown in Supplementary Table S1. Calibrations of R were performed at regular intervals over several years to characterize the small drift, which is common in artefact standards: R=1000051260±800 Ω on 27/4/2011, with a drift correction of 11Ω per day. The maximum drift correction applied for pump measurements before and after the calibration date was ≈0.5 p.p.m. The random uncertainty UR is dominated by the Johnson current noise in R≈4 fA/√Hz. Thus, 1 p.p.m. resolution of IP≈100 pA requires averaging times of the order of 1 h.
Additional information
How to cite this article: Giblin, S.P. et al. Towards a quantum representation of the ampere using single electron pumps. Nat. Commun. 3:930 doi: 10.1038/ncomms1935 (2012).
References
Zimmerman, N. M. A primer on electrical units in the systeme international. Am. J. Phys. 66, 324–331 (1998).
Kibble, B. P. Present state of the electrical units In Science, Measurement and Technology, IEE Proceedings A Vol 138 (IET, 1991) pp. 187–197.
Milton, M. J. T., Williams, J. M. & Forbes, A. B. The quantum metrology triangle and the redefinition of the SI ampere and kilogram; analysis of a reduced set of observational equations. Metrologia 47, 279–286 (2010).
Zimmerman, N. M. & Keller, M. W. Electrical metrology with single electrons. Meas. Sci. Technol. 14, 1237–1242 (2003).
Keller, M. W., Martinis, J. M., Zimmerman, N. M. & Steinbach, A. H. Accuracy of electron counting using a 7-junction electron pump. Appl. Phys. Lett. 69, 1804–1806 (1996).
Keller, M. W., Eichenberger, A. L., Martinis, J. M. & Zimmerman, N. M. A capacitance standard based on counting electrons. Science 285, 1706–1709 (1999).
Shilton, J. M. et al. High-frequency single-electron transport in a quasi-one-dimensional GaAs channel induced by surface acoustic waves. J. Phys. 8, L531–L539 (1996).
Pekola, J. P. et al. Hybrid single-electron transistor as a source of quantized electric current. Nat. Phys. 4, 120–124 (2007).
Maisi, V. F., Pashkin, Y. A., Kafanov, S., Tsai, J. S. & Pekola, J. P. Parallel pumping of electrons. N. J. Phys. 11, 113057 (2009).
Janssen, T.J.B.M. & Hartland, A. Accuracy of quantized single-electron current in a one-dimensional channel. Physica B 284, 1790–1791 (2000).
Zimmerman, N. M., Hourdakis, E., Ono, Y., Fujiwara, A. & Takahashi, Y. Error mechanisms and rates in tunable-barrier single-electron turnstiles and charge-coupled devices. J. Appl. Phys. 96, 5254–5266 (2004).
Fujiwara, A., Nishiguchi, K. & Ono, Y. Nanoampere charge pump by single-electron ratchet using silicon nanowire metal-oxide-semiconductor field-effect transistor. Appl. Phys. Lett. 92, 042102 (2008).
Kashcheyevs, V. & Kaestner, B. Universal decay cascade model for dynamic quantum dot initialization. Phys. Rev. Lett. 104, 186805 (2010).
Wright, S. J. et al. Enhanced current quantization in high-frequency electron pumps in a perpendicular magnetic field. Phys. Rev. B 78, 233311 (2008).
Kaestner, B. et al. Single-parameter quantized charge pumping in high magnetic fields. Appl. Phys. Lett. 94, 012106 (2009).
Leicht, C. et al. Generation of energy selective excitations in quantum hall edge states. Semiconductor Sci. Technol. 26, 055010 (2011).
Wright, S. J. et al. Single- and few-electron dynamic quantum dots in a perpendicular magnetic field. J. Appl. Phys. 109, 102422 (2011).
Fletcher, J. D. et al. Stabilisation of single-electron pumps by high magnetic fields (2011)Preprint at arXiv:1107.4560.
Blumenthal, M. D. et al. Gigahertz quantized charge pumping. Nat. Phys. 3, 343–347 (2007).
Giblin, S. P. et al. An accurate high-speed single-electron quantum dot pump. N. J. Phys. 12, 073013 (2010).
Kataoka, M. et al. Tunable nonadiabatic excitation in a single-electron quantum dot. Phys. Rev. Lett. 106, 126801 (2011).
Mohr, P. J., Taylor, B. N. & Newell, D. B. CODATA recommended values of the fundamental physical constants: 2006. Rev. Mod. Phys. 80, 633–730 (2008).
Camarota, B. et al. Electron counting capacitance standard with an improved five-junction Rpump. Metrologia 49, 8–14 (2012).
Keller, M. W., Zimmerman, N. M. & Eichenberger, A. L. Uncertainty budget for the NIST electron counting capacitance standard, ECCS-1. Metrologia 44, 505–512 (2007).
Piquemal, F. & Geneves, G. Argument for a direct realization of the quantum metrological triangle. Metrologia 37, 207–211 (2000).
Rietveld, G. et al. 1: 30000 cryogenic current comparator with optimum SQUID readout. Instrum. Meas., IEEE Trans. 52, 621–625 (2003).
Feltin, N., Devoille, L., Piquemal, F., Lotkhov, S. V. & Zorin, A. B. Progress in measurements of a single-electron pump by means of a CCC. Instrum. Meas., IEEE Trans. 52, 599–603 (2003).
Steck, B. et al. Characterization and metrological investigation of an R-pump with driving frequencies up to 100 Mhz. Metrologia 45, 482–491 (2008).
Keller, M. W. Current status of the quantum metrology triangle. Metrologia 45, 102–109 (2008).
Fricke, L. et al. Quantized current source with mesoscopic feedback. Phys. Rev. B 83, 193306 (2011).
Yamahata, G., Nishiguchi, K. & Fujiwara, A. Accuracy evaluation of single-electron shuttle transfer in si nanowire metal-oxide-semiconductor field-effect transistors. Appl. Phys. Lett. 98, 222104 (2011).
Acknowledgements
This research was supported by the UK Department for Business, Innovation and Skills, the European Metrology Research Programme (grant no. 217257) and the UK EPSRC.
Author information
Authors and Affiliations
Contributions
S.P.G. designed and calibrated the measurement system, M.K. designed the pulse drive technique, J.D.F. designed the measurement probe, S.P.G. and M.K. performed experiments and analysed data, M.K. designed the sample, P.S. performed and supervised sample fabrication, D.A.R. and I.F. provided GaAs wafers, J.P.G. and G.A.C.J. performed electron beam lithography, T.J.B.M.J. contributed project leadership and supervision, and S.P.G. wrote the paper with extensive comments/revisions contributed by M.K. and J.D.F.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing financial interests.
Supplementary information
Supplementary Information
Supplementary Figure S1 and Supplementary Table S1 (PDF 103 kb)
Rights and permissions
About this article
Cite this article
Giblin, S., Kataoka, M., Fletcher, J. et al. Towards a quantum representation of the ampere using single electron pumps. Nat Commun 3, 930 (2012). https://doi.org/10.1038/ncomms1935
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/ncomms1935
This article is cited by
-
Time-resolved Coulomb collision of single electrons
Nature Nanotechnology (2023)
-
A macroscopic mass from quantum mechanics in an integrated approach
Communications Physics (2022)
-
Biodegradable films functionalized with Moringa oleifera applied in food packaging
Iranian Polymer Journal (2021)
-
Continuous-variable tomography of solitary electrons
Nature Communications (2019)
-
High-accuracy current generation in the nanoampere regime from a silicon single-trap electron pump
Scientific Reports (2017)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.