Abstract
We report a case of severe erosion and cellulitis at the base of the penis as a result of vacuum suction device constriction bands left on for 4 hours in a spinal cord injured patient with paraplegia and hypesthesia of the genital area. All patients using vacuum suction devices need to be properly educated regarding usage and risks with adequate follow up; patients with hypesthesias and spinal cord injuries need information specifically related to their decreased or absent level of sensation. Only two out of seven vacuum suction device brochures reviewed warn of the risk to patients with decreased sensation in the penis, but none specifically address usage or risks to men with spinal cord injuries.
Similar content being viewed by others
Article PDF
References
Witherington R (1989) Vacuum constriction device for management of erectile impotence. J Urol 141:320–322.
Witherington R (1988) Suction device therapy in the mangement of erectile impotence. Urol Clin North Am 15: 123.
Salvatore F, Sharman G, Hellstrom W (1991) Vacuum constriction devices and the clinical urologist: an informed selection. J Urol 38: 323–327.
Lewis J, Sidi A, Reddy P (1991) A way to help your patients who use vacuum suction devices. Contemp Urol December: 15–24.
Mission Pharmacal Company (1991) Mission Vacuum Erection Device (VEDR) User's Guide, San Antonio, Texas.
Dacomed Corporation (1991) Dacomed Corporation Catalystâ„¢ Vacuum System. Minneapolis, MN.
Mentor Corporation (1991) Mentor Vacuum Suction Device (Touchâ„¢, Pistonâ„¢, Responseâ„¢).Santa Barbara, CA.
Osbon Medical Systems Ltd (1991) Guide to the Osbon Technique: How to use ErecaidR System Classic, Augusta, GA.
Post-T-Vacâ„¢ Inc (1991) Pos-T-Vacâ„¢ System. Dodge City, KS.
Moul J, McLeod D (1989) Negative pressure devices in the explanted penile prosthesis population. J Urol 142: 729–731.
NuMedTec Inc (1991) E/P System, Westchester, PA.
Smith Collins Pharmaceutical Inc (1992) SCP 2000/3000, Vernon Hills, IL.
Meinhardt W, Kropman R, Lycklama A, Nijeholt A, Zwartendijk J (1990) Skin necrosis caused by use of negative pressure device for erectile impotence. J Urol 144: 983.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
LeRoy, S., Pryor, J. Severe penile erosion after use of a vacuum suction device for management of erectile dysfunction in a spinal cord injured patient. Case report. Spinal Cord 32, 120–123 (1994). https://doi.org/10.1038/sc.1994.22
Issue Date: