Main

Glutamate and γ-aminobutyric acid (GABA) are the major excitatory and inhibitory neurotransmitters, respectively, in the brain and are likely involved in nearly all signal processing functions of the central nervous system (CNS) and altered in many CNS diseases1,2,3,4. Magnetic resonance imaging (MRI) is a noninvasive imaging technique that provides exquisite structural detail, but current MRI methods are not capable of imaging the distribution of neurotransmitters in brain (Supplementary Note).

Proton magnetic resonance spectroscopy (1H MRS) can detect several neurotransmitter signature groups using a variety of techniques5,6,7,8,9,10. However, 1H MRS techniques require long acquisition times and have poor spatial resolution. Here we describe a new MRI technique for imaging glutamate (R.R., K.C., M.H., A.S., F.K. and H.H., unpublished data) based on its chemical exchange saturation transfer (CEST) effect11,12,13,14,15. GluCEST provides markedly higher spatial and temporal resolution compared with 1H MRS. The CEST effect from amide and hydroxyl protons from different amino acids, proteins and other molecules12,16,17 have previously been exploited to measure pH in vivo18,19, glycogen in the liver20, glycosaminoglycans in cartilage21, gene expression in vivo22 and myo-inositol in brain23, but no prior studies have used the CEST effect of amine protons to image glutamate in vivo.

Here the pH and concentration dependence of GluCEST is first demonstrated in vitro at 37 °C. Second, the feasibility of measuring GluCEST in vivo from rat brain with and without focal ischemia is shown. Third, changes in GluCEST and glutamate 1H MRS in response to exogenous injection of glutamate in a rat brain tumor model are compared. Finally, the feasibility of mapping the glutamate signal from healthy human brain in vivo is demonstrated at 7 Tesla (7T), an ultrahigh-field whole-body imaging scanner. The potential contributions to GluCEST from other brain metabolites, as well as the advantages and shortcomings of this approach, are also discussed.

Results

Phantom studies

The z-spectra of 10 mM glutamate solutions at varying pHs (phantoms) demonstrate that at lower pHs (pH < 6.0), the CEST peak of glutamate is sharper and centered around 3 p.p.m. higher frequency from the bulk water resonance (absolute chemical shift, 7.7 p.p.m.) (Fig. 1a). The z-spectral asymmetry plots (Fig. 1b) gave a broad CEST effect ranging from 1 to 4 p.p.m. relative to bulk water resonance. Whereas the GluCEST asymmetry plots from low pH solutions showed clear peaks centered at 3.00 p.p.m., the peak position of asymmetry plots from higher pH solutions (pH >6.0) gradually moved toward that of water resonance. At pH 7, the maximum of the asymmetry plot occurred at 1.2 p.p.m., and its amplitude was 100% higher than at 3 p.p.m. This is the well known phenomenon of intermediate- to fast-exchange (k ≥ Δω)-mediated chemical shift averaging. Despite the shift in peak position at pH 7, all of our studies reported here used 3 p.p.m. as the GluCEST resonance saturation frequency to maintain consistency.

Figure 1: GluCEST: pH dependence and sensitivity.
figure 1

(a) CEST z-spectra at 7T of 10 mM glutamate at varying pHs at 37 °C show the CEST effect at 3 p.p.m. downfield to bulk water resonance. Msat and M0 represent bulk water signals acquired with and without selective RF saturation, respectively. (b) GluCEST asymmetry curves from different pH solutions corresponding to a. The overlay of simulated GluCEST (dotted line) on the curve corresponding to pH 7 is also shown. (c) Dependence of GluCEST on pH 3–8. The fitted line (dashed line generated with the equation shown) shows the linear dependence of GluCEST on pH in the physiological range (pH 6.0–7.4). (d) Image from a phantom of 10 mM glutamate (pH 7.0, 37 °C) prepared in PBS, with the localization voxel indicated by the square. (e) 1H MRS point-resolved spectroscopy (PRESS) water-suppressed spectrum (repetition time (TR) = 20 s, echo time (TE) = 16 ms, four averages) from the single voxel shown in d. The three resonances correspond to two −CH2 groups and one −CH group of glutamate. (f) Water 1H resonance spectra obtained while saturating at ± 3 p.p.m., as well as their difference spectrum (blue line). (g) Rescaled difference spectrum from f. The difference spectrum represents the 3 p.p.m. point on the CEST asymmetry spectrum (CESTasym). The ratio of the CEST difference spectrum to glutamate –CH2 resonance at 2.3 p.p.m. is 700. MR, magnetic resonance; AU, arbitrary unit.

Because of non–steady-state experimental conditions, we analyzed the sensitivity of GluCEST by solving numerical simulations of full Bloch-McConnell equations24 using the experimental parameters from these studies. We found an excellent agreement between the experimental z-spectral asymmetry curves and numerically simulated ones. Because the intracellular pH in the brains of normal healthy volunteers is reported to be 7.0 (ref. 25), the GluCEST contrast at this pH is the most relevant for in vivo imaging. On the basis of the fitted z-spectral asymmetry curve from simulations (Fig. 1b), the amine-proton exchange rate (k) in a 10 mM glutamate solution (pH 7.0) at 37 °C was estimated to be in the range of 5,500 ± 500 s−1. This is comparable to the chemical shift difference (Δω) at 7T (Δω = 5,700 rad s−1); thus, GluCEST imaging is feasible at 7T, as well as at higher field strengths.

For the experimental saturation parameters used here, GluCEST increased gradually from pH 3 to pH 6 and then sharply decreased from pH 6 to 8. The largest GluCEST observed at pH 6 is probably caused by the estimated reduced exchange rate of amine protons with the bulk water. GluCEST varies linearly with pH in the physiological range (pH 6.0–7.4) (Fig. 1c). There is a 700-fold sensitivity advantage of GluCEST over conventional 1H MRS at physiological concentration and temperature (Fig. 1d–g). This sensitivity amplification should make it feasible to detect relatively small changes in glutamate levels at high resolution (voxel size of 0.02 cc).

The GluCEST effect is linearly proportional to the glutamate concentration in the physiological range at pH 7 (Fig. 2a). With the imaging parameters used here, an 0.6% CEST effect increase per mM glutamate was observed (Fig. 2b). The GluCEST effect increased with peak radiofrequency (RF) field (B1) amplitude and saturation pulse duration, demonstrating that it is possible to obtain higher GluCEST effects (>1% per mM glutamate) (Fig. 2c). Whereas it is possible to achieve a higher GluCEST sensitivity by using higher B1, the imaging parameters chosen here were constrained by the exchange rate differences between phantoms and in vivo tissue, as well as RF-specific absorption rate (SAR) issues in human studies at 7T. As SAR is proportional to the B12, and the square of the static magnetic field (B02), the allowed B1 amplitude at 7T is limited for human imaging.

Figure 2: GluCEST images at 7T of a phantom consisting of test tubes with different concentrations of glutamate (pH 7.0) immersed in a beaker containing PBS.
figure 2

All the experiments were performed at 37 °C. (a) GluCEST contrast color-coded on the original CEST image (3 p.p.m.), acquired with application of saturation pulse train with B1rms = 155 Hz (3.6 μT) for 2 s. Color bar represents percentage GluCEST contrast. (b) Linear dependence (R2 = 0.97) of GluCEST effect on glutamate (Glu) concentration with a slope of 0.6% per mM glutamate. (c) GluCEST dependence on B1rms and duration of the saturation pulse. (d,e) CEST images of a phantom consisting of test tubes immersed in a beaker containing PBS, with solution of the different metabolites (pH 7) at their physiological concentrations. Glu, 10 mM glutamate; GABA, 2 mM GABA; NAA, 10 mM N-acetyl aspartate; Gln, 2 mM glutamine; Asp, 2 mM aspartate; 2 mM taurine; Cr, 6 mM creatine and MI, 10 mM myo-inositol. The CEST contrast color-coded on the original CEST images (3 p.p.m.), showing substantial CEST contrast from glutamate (6%), with 1% from GABA, <0.5% from creatine and no contrast from other metabolites. Color bar represents percentage CEST contrast. The error bars represent s.d. from repeated experiments.

Potential contributions from other brain metabolites

The major metabolites in the brain visible with 1H MRS at physiological concentrations are N-acetyl aspartate, choline, myo-inositol, creatine, glutamine, taurine, aspartate, glutamate and GABA. After investigating the CEST properties from each of these metabolites in phantoms, we found that, with the experimental parameters used here, except for a small contribution from GABA and creatine, all other metabolites contributed negligible CEST effects to GluCEST (Fig. 2d,e). Of these metabolites, N-acetyl aspartate has an amide proton, but at neutral pH it did not show any CEST effect. Choline does not have any exchangeable protons. Myo-inositol has exchangeable hydroxyl protons and showed a CEST effect at ≤1 p.p.m., which did not extend beyond 2 p.p.m. at neutral pH. Creatine has amine protons that showed a CEST effect around 1.8 p.p.m. at physiological pH. However, its contribution to the GluCEST effect at 3 p.p.m. was also not appreciable (<0.5%). No contributions to GluCEST were observed from aspartate, glutamine and taurine at their physiological concentrations under the experimental conditions used, possibly because of unfavorable exchange rates and differences in the chemical shifts of their amine protons.

Rat studies

In healthy rat brain, the gray matter and white matter were clearly separated in the GluCEST, with a contrast ratio of 1.6 (Fig. 3a–d). The glutamate amine exchange rate measured in healthy rat brain was 2,000 ± 500 s−1 (Supplementary Methods), which is well within the slow to intermediate exchange regime at 7T.

Figure 3: GluCEST mapping of healthy and ischemic rat brain.
figure 3

(ad) Anatomic image, GluCEST map, local static magnetic field (B0) and radiofrequency field (B1) maps of healthy rat brain slice, respectively. Clear differences of GluCEST contrast in gray matter and white matter regions can be seen. (e) MCAO rat brain anatomic proton image. (f,g) GluCEST maps acquired at 1 h and 4.5 h after the induction of stroke. (h) The GluCEST contrast versus time after MCAO at regions of interest within the rectangular areas shown in g. (i) The GluCEST asymmetry plots from the contralateral side (blue curve) and ipsilateral side (red curve) at 4.5 h after occlusion. (j) The contralateral and ipsilateral sides of MCAO rat brain data acquired at 4.7T at 4 h after occlusion and 1 h post-mortem. (Panel j reprinted from ref. 18.). The error bars represent s.d. from repeated experiments.

GluCEST obtained from a rat brain following middle cerebral artery occlusion (MCAO) stroke showed significant differences between ipsilateral and contralateral sides (Fig. 3e–h). The ipsilateral side demonstrated an 100% increase in GluCEST at 4.5 h after MCAO (Fig. 3h). The z-spectral asymmetry plots from the MCAO brain reported here showed an elevation of CEST, whereas z-spectra from a previously reported MCAO study in the rat brain18 have shown decreased amide proton CEST effects from the ipsilateral side compared with the contralateral side (Fig. 3i,j). In addition, amide proton CEST asymmetry plots have shown negative values at 3 p.p.m.18, whereas our GluCEST showed positive values. This is because of key differences in the amide proton CEST and GluCEST experimental conditions. Because of low exchange rates (30 s−1) in amide proton CEST, a lower saturation pulse amplitude (root mean square B1 (B1rms) = 50 Hz (1.2 μT)) and longer duration (6 s) were used in the earlier study. However, owing to the higher glutamate exchange rate, higher B1rms and shorter duration were used in this study (B1rms = 155 Hz (3.6 μT) and 1 s).

Literature studies have shown that in the MCAO model of ischemia, extracellular glutamate levels (1–2 μM basal value)26 increase substantially to 0.2 mM (ref. 27). However, the total glutamate concentration (extracellular and intracellular) may remain unaltered. In addition, although extracellular concentrations (in μM) of other molecules such as GABA, aspartate, adenosine and taurine also increase28, their concentration changes are too small to account for the observed GluCEST effects.

In the MCAO model, it is also well established that the magnitude of pH decrease expected over a 4-h period is 0.5 units18. In addition to pH, other factors that potentially affect GluCEST during stroke are changes to the water content, water spin-spin relaxation time (T2) and magnetization transfer ratio (MTR). It has been reported that in the MCAO model, the T2 of water is elevated by 20 to 40%, MTR is reduced by 20% and total water content may be changed by 2 to 4%18,29. These changes in T2, MTR and water content are expected to account for only minor changes in GluCEST.

In polypeptide solutions in PBS, the CEST effect from amine and amide protons has been shown to decrease with decreasing pH in the pH 7.3–5.0 range30. In contrast, GluCEST increased here by almost 100% during MCAO ischemia. This increase in GluCEST is consistent with the pH-dependent GluCEST behavior observed in glutamate phantoms at physiological concentrations and temperature, and it suggests that the elevated GluCEST in the ipsilateral side of MCAO model is predominantly because of lower pH.

In healthy brain, glutamate does not cross the blood-brain barrier (BBB); however, the compromised BBB in a tumor model provides a means of demonstrating glutamate modulation, where GluCEST should change approximately linearly with the concentration of glutamate. To demonstrate that the GluCEST measured in vivo is specific to glutamate, we performed GluCEST and 1H MRS studies in a rat brain with a tumor before and after the intravenous injection of glutamate solution (Fig. 4a–f). GluCEST maps obtained at three time points after injection showed that there was a notable increase in GluCEST in the tumor, with a concomitant similar elevation in the 1H MRS of glutamate.

Figure 4: GluCEST images and 1H MRS of rat brain with a tumor, before and after injection of glutamate.
figure 4

(a) Rat brain anatomic proton image demonstrating the tumor and a rectangular region of interest. (b) B0 map of the rat brain, which shows about ± 0.5 p.p.m. variation. Color bar is in units of p.p.m. (c) B1 map (the ratio of B1/B1ref, unitless) of the rat brain showing a fairly uniform B1 field. (d) GluCEST maps of the rat brain acquired before injection (BSL) and at 1, 1.5 and 2.0 h after the injection of glutamate solution. Color bars represent percentage GluCEST contrast. (e) Stacked stimulated echo acquisition mode (STEAM) localized 1H MRS data (in arbitrary MR units) acquired at baseline and at 2 h after the glutamate (Glu) injection (arrows). (f) Time course of GluCEST and the 2.3 p.p.m. glutamate peak integral normalized with the values of pre-injection from regions of interest within the rectangular areas shown in a.

In vivo human brain study

A GluCEST map obtained from a healthy human brain (Fig. 5) showed a gray matter and white matter distribution pattern similar to a published positron emission tomography (PET) map of the metabotropic glutamate receptor subtype 5 (mGluR5) (Fig. 5c and ref. 31). Concentrations of glutamate derived from the 1H MRS data (5.9 ± 0.5 mM for white matter and 9.4 ± 1.4 mM for gray matter) were also consistent with the GluCEST from the same regions of interest (ROIs) (Fig. 5a,b,f).

Figure 5: GluCEST imaging and 1H MRS from a healthy human brain acquired at 7T.
figure 5

(a) Anatomic proton image of the axial slice. (b) B1 and B0 corrected GluCEST contrast map. Color bar represents percentage GluCEST contrast. (c) Map of distribution volumes (DVs) of metabotropic glutamate receptor subtype 5 from a PET image (reprinted from ref. 31 with permission of the Society of Nuclear Medicine). (d) B0 map and (e) B1 map corresponding to the slice in a. (f) 1H MRS spectra obtained from regions of interest of gray matter and white matter shown in a. These spectra (in arbitrary MR units) show higher amplitude of glutamate −CH2 resonance (2.3 p.p.m.) and −CH resonance (3.75 p.p.m.) in gray matter than that in white matter. (g) Saturation pulse duration dependence of GluCEST data from human brain. The GluCEST reaches a maximum at a saturation duration of 1 s and decreases with further increase in duration. (h,i) The z-spectra and corresponding asymmetry plots from human gray matter and white matter regions. The GluCEST at 3 p.p.m. (dotted line in i) is 11% from gray matter and 7% from white matter. Glu, glutamate; GM, gray matter; WM, white matter.

In human brain, the gray matter and white matter GluCEST contrast was the highest at a saturation pulse length of 1 s (Fig. 5g). The z-spectra and CEST asymmetry curves from gray matter and white matter regions of the brain were rather broad and showed a maximum CEST contrast at 2 p.p.m. (Fig. 5i,j). The broadness in asymmetry plots may be partly due to (i) the glutamate exchange–induced chemical shift averaging effect that shifts the line toward water resonance and (ii) an additional potential contribution from creatine32. Similar results were observed in three volunteers. The intrasubject coefficient of variation of measured GluCEST was <5%.

The GluCEST maps obtained in healthy human brain reflected the known regional variation of physiological concentrations of glutamate33,34. Because 1H MRS data are highly specific to the concentration of glutamate, the observed strong correlation between GluCEST ratio from gray matter and white matter ROIs (1.6 ± 0.2) with the glutamate concentration ratio from 1H MRS data in the same ROIs (1.6 ± 0.1) provided additional confirmation that the observed GluCEST in the human brain is predominantly due to glutamate. We estimate that 70–75% of the observed GluCEST was from glutamate (Supplementary Table 1).

Discussion

These findings provide what is to our knowledge the first evidence that GluCEST can be used for noninvasive imaging of glutamate concentrations, with excellent spatial and temporal resolution. In addition to its role as the predominant excitatory neurotransmitter in the brain, glutamate also serves as a metabolic intermediate, and it is likely that GluCEST detects both of these pools.

Although the gray matter to white matter GluCEST contrast ratio in humans, as well as in healthy rat brains, is very similar (1.6 ± 0.2), the GluCEST magnitude in the rat brain is higher than that in the human brain because of higher strength magnetic fields (9.4T in rats versus 7T in humans), continuous saturation pulses used on rats (constrained by the scanner software) and lower direct water saturation at 9.4T in rats compared with 7T in humans. Because of the fast exchange rate of glutamate amine protons, the slow to intermediate exchange rate condition is not fulfilled at lower-field whole-body clinical scanners (≤3T), and, hence, these studies are expected to have poor sensitivity at these fields.

Potential contributions of MTR asymmetry to GluCEST, caused by the bound pool of water from rigid macromolecules in biological tissues, cannot be excluded. MTR asymmetry effects35, if any, would compete with the GluCEST effect and would lead to an underestimation of actual GluCEST levels. Indeed, the simulated MTR asymmetry (−0.14% in gray matter and −0.22% in white matter) (Supplementary Table 1 and Supplementary Discussion) indicates that its contribution to the observed GluCEST is likely to be negligibly small.

Here SAR issues at 7T limit the observed GluCEST sensitivity in humans. However, saturation pulse power, duration, pulse shapes and readout sequences, in conjunction with multichannel RF coils, could be further optimized to enhance GluCEST effects within FDA guidelines for SAR deposition. The glutamate contribution to GluCEST may have slight variations per site, depending on the RF coil or static field strength used, and future work should evaluate these contributions for specific experimental conditions.

Studies at ultrahigh fields beyond 7T would further enhance the sensitivity of GluCEST, albeit with additional challenges in dealing with SAR deposition. For proof-of-principle purposes, we used only a single slice with a two-dimensional imaging sequence. However, it would be relatively straightforward to implement a three-dimensional acquisition of GluCEST maps.

In summary, GluCEST detection is feasible in the human brain at ultrahigh field (7T) without exceeding the allowed limits on SAR. About 70–75% of the observed CEST is from glutamate, with the remaining 25–30% coming from creatine, GABA and other macromolecules. Despite small contamination from other molecules in vivo, the GluCEST method can be used to study relative changes in glutamate through endogenous or exogenous glutamate modulation in vivo. This method provides noninvasive, nonradioactive and high spatial and temporal resolution imaging of relative changes in glutamate for use in both human subjects and preclinical models. Further, the exquisite sensitivity of GluCEST to changes in pH could potentially be used as a pH marker in evolving stroke and tumors. Future studies using this approach may provide new insights into glutamate function and demonstrate its potential as a biomarker for the diagnosis and treatment of CNS disorders.

Methods

MRI scans.

All the phantom and human MRI studies were performed in a 7T Siemens whole body MRI scanner (Siemens Medical Systems) with a circularly polarized RF coil. All phantoms were prepared in PBS, and all the experiments were performed at 37 °C. Rat experiments were performed in a 9.4T Varian scanner (Agilent Technologies).

Phantom imaging.

A new pulse sequence was written to use a frequency selective saturation pulse followed by a segmented RF spoiled gradient echo (GRE) readout sequence. The imaging parameters were as follows: slice thickness = 10 mm, flip angle = 10°, TR = 5.6 ms, TE = 2.7 ms, field of view (FOV) = 120 × 120 mm2, matrix size = 192 × 192, with one saturation pulse and 64 segments acquired every 15 s. CEST images were collected using a saturation pulse at a B1rms of 155 Hz (3.6 μT) and 2-s duration and frequencies from −5 to 5 p.p.m. in steps of 0.2 p.p.m.

Rat experiments.

All rat experiments were performed according to an approved University of Pennsylvania Institutional Animal Care and Use Committee protocol. For healthy and MCAO studies, Sprague-Dawley male rats (265–315 g, Charles River Laboratories) were used, and, for tumor model studies, female Fisher rats (130–150 g, Charles River Laboratories) were used. Both MCAO and tumor models were prepared as described in the Supplementary Methods.

Rat imaging and spectroscopy.

Healthy (n = 2), MCAO (n = 3) and tumor-bearing rats (n = 3) were imaged using a 35-mm-diameter quadrature RF coil (m2m Imaging Corp.). Animals were kept under anesthesia (1.5% isoflurane in 1 l min−1 oxygen) and kept warm with the warm air from a heater. Respiration and body temperature were monitored using an MRI-compatible small animal monitoring system (SA Instruments). CEST imaging was performed using a frequency selective continuous wave saturation pulse followed by a segmented GRE readout sequence. The imaging parameters were FOV = 35 × 35 mm2, slice thickness = 2 mm, flip angle = 15°, TR = 6.2 ms, TE = 2.9 ms, matrix size = 128 × 128 and number of averages = 2, with one saturation pulse and 32 segments acquired every 4 s. CEST images were collected using a 1-s-long rectangular saturation pulse at B1rms of 250 Hz (5.9 μT) at multiple frequencies in the range −5 to 5 p.p.m. with a step size of 0.2 p.p.m. Data for B1 and B0 maps were also acquired. In vivo exchange rate in healthy rat brain at 9.4T was measured using a previously described method14. In tumor-bearing rats, along with GluCEST data, water suppressed STEAM single voxel spectra were obtained with the following parameters: voxel size = 11 × 8 × 5 mm3, spectral width = 4 kHz, number of points = 2048, average = 128, TE = 8 ms and TR = 6 s. After collecting the baseline CEST map and 1H MRS, the rats were injected with 2.5 ml of 100 mM glutamate solution through the tail vein. CEST and 1H MRS data were gathered periodically for 2 h after the injection.

Human studies.

The studies were conducted under an approved Institutional Review Board protocol of the University of Pennsylvania. GluCEST imaging and z-spectrum acquisitions of human brain at 7T were performed on three normal volunteers (male, ages 27–35 years). Informed consent from each volunteer was obtained after explaining the study protocol. Brain CEST imaging was performed with the application of the same imaging protocol as described for the phantoms except with a larger FOV (240 × 240 mm2), averages (4) and matrix size (128 × 128). Original CEST images and GluCEST contrast maps were corrected for B0 and B1 in homogeneities (Supplementary Methods). Water suppressed 1H MRS were obtained from gray matter and white matter regions using the standard point resolved spectroscopy (PRESS)36 localization technique with the following parameters: voxel size = 10 × 40 × 10 mm3, spectral width = 4 kHz, number of points = 2,048, average = 100, TE = 16 ms, and TR = 3 s.

Data processing.

All image processing and data analysis was performed using in-house written programs in MATLAB (version 7.5, R2007b). Acquired images were corrected for B0 and B1 inhomogeneities, and z-spectra were obtained from these images by plotting the normalized image intensity as a function of resonance offset of saturation pulse for each sample. CEST contrast was calculated with the following equation37:

where Msat (±Δω) represent the magnetizations obtained with saturation at a positive or negative Δω offset from the water resonance. CESTasym contrast at 3 p.p.m. is referred to as GluCEST. 1H MRS spectra were fitted using nonlinear least-squares methods with Gaussian functions. Peak integrals were calculated and normalized with a non–water-suppressed proton signal for calculating glutamate concentrations in vivo.

Additional methods.

Detailed methodology is described in the Supplementary Methods.