Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Article
  • Published:

Validation of Omron HEM-9200T, a home blood pressure monitoring device for the upper arm, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 protocol

Abstract

The performance of Omron HEM-9200T for monitoring blood pressure (BP) in the upper arm was validated in accordance with the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 protocol. The device was assessed by using it on 87 participants who fulfilled the inclusion criteria involving the ranges of arm circumference and systolic and diastolic BP provided by the protocol. Validation and data analysis were performed according to the protocol. In the ANSI/AAMI/ISO 81060-2:2013 validation procedure (criterion 1), the mean ± standard deviation of the differences between the test device and reference BP was −0.1 ± 5.06/1.2 ± 5.8 mmHg (systolic/diastolic). The mean differences between the two observers and Omron HEN-9200T were −0.1 ± 3.82 mmHg for systolic BP and 1.2 ± 5.34 mmHg for diastolic BP, fulfilling criterion 2 with an SD of ≤6.91 for SBP and ≤6.87 for DBP. These two ANSI/AAMI/ISO criteria were fulfilled.

The Omron HEM-9200T BP monitor fulfilled the requirements of the ANSI/AAMI/ISO validation standard and can be recommended for BP measurements at home in the general population.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Fig. 1: Bland-Altman plots.

Similar content being viewed by others

References

  1. Association for the Advancement of Medical Instrumentation American National Standard. ANSI/AAMI/ISO 81060-2:2013 non-invasive sphygmomanometers – part 2: clinical investigation of automated measurement type. http://my.aami.org/aamiresources/previewfiles/8106002_1306_preview.pdf#search=%27American+National+Standard.+ANSI%2FAAMI%2FISO+810602%3A2013+Noninvasive%27. Accessed 15 Sep 2020.

  2. Stergiou GS, Alpert B, Mieke S, Asmar R, Atkins N, Eckert S, et al. A universal standard for the validation of blood pressure measuring devices: association for the advancement of medical instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. Hypertension. 2018;71:368–74.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hakuo Takahashi.

Ethics declarations

Conflict of interest

KS and NY are employed by Omron Healthcare Co., Ltd, the manufacturer of Omron HEM-9200T. The authors report no other conflicts of interest in this work.

Additional information

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Takahashi, H., Saito, K. & Yakura, N. Validation of Omron HEM-9200T, a home blood pressure monitoring device for the upper arm, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 protocol. J Hum Hypertens 36, 416–419 (2022). https://doi.org/10.1038/s41371-021-00536-1

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/s41371-021-00536-1

This article is cited by

Search

Quick links