Table 1 m-PTAS sensor detection performance.
From: A Portable Smartphone-linked Device for Direct, Rapid and Chemical-Free Hemoglobin Assay
Analysis Category | Specimen Type | Specimen[Hb] (g/dL) | m-PTAS Sensor | Limit of Blank (g/dL) | |||
---|---|---|---|---|---|---|---|
Mean (rad) | SD (rad) | Measurement | |||||
Limit of Blank | Plasma | 0.0000 | 0.0469 | 0.2811 | 60 | 0.0039 | |
Analysis Category | Specimen Type | Specimen[Hb] (g/dL) | m-PTAS Sensor | ||||
Mean (g/dL) | SD (g/dL) | Measurement | Limit of Detection (g/dL) | ||||
Limit of Detection | Diluted Blood Specimen | 0.0316 | 0.0371 | 0.0263 | 60 | 0.0471 | |
Analysis Category | Specimen Type | Specimen[Hb] (g/dL) | m-PTAS Sensor | ||||
Mean (g/dL) | SD (g/dL) | CV (%) | Measurement | Limit of Quantitation (g/dL) | |||
Limit of Quantitation | Diluted Blood Specimen | 0.0475 | 0.0492 | 0.0205 | 41.74 | 10 | 0.1069 |
0.0950 | 0.0877 | 0.0237 | 27.04 | 10 | |||
0.1188 | 0.1229 | 0.0201 | 16.36 | 10 | |||
0.1900 | 0.1872 | 0.0224 | 11.98 | 10 | |||
0.2375 | 0.2529 | 0.0265 | 10.49 | 10 |